The range

.

The prosthesis is an anatomic right and left prosthesis with a 135 ° sloping angle and an 10 ° anteversion angle.

The titanium alloy metaphysiary part drawing (TA16V4) provides a suited metaphysiary filling, as, when installed, it comes to rest both on the lateral face of the greater trochanter and on the front and rear side of the metaphysis ; its surface has horizontal fissures in order to prevent sinking. Its rough surface is covered with a layer of 70 micron hydroxyapatite intended to facilitate its secondary fastening.

The shaft part is also titanium ; it is smooth, without hydroxyapatite, rather short and is available in 7 sizes : it is used for temporary centring pending the final metaphysiary anchoring.

70 combinations are possible, matching most standard femurs.

The fastening of both parts has been the subject of numerous studies. The parts are cylindrically inserted in one another and screwed.

A finished element study has been conducted prior to the use of the prosthesis. The purpose of the study was to study the behaviour of both fastened parts in terms of warping and constraints, mainly at the shaft / metaphysis unit. Various components were studied: alumina ball, metaphysis parts, shaft parts, cortical bones, upper spongy bone and lower spongy bone. All 6 components were subjected to a 2 000 N stress applied to the alumina ball (Pauwels' diagram). The results show that the level of stress is very low. The maximum figures (15.4 daN/mm2) were found on the metaphysis part, in the area connecting the stem holding the ball and the massive section. With the assembly of both parts, the maximum stress value reached 8 daN / mm2, in spots only. It was found that the maximum stress values (15 daN / mm2) were not dangerous for the metaphysis and the shaft since the value is much lower than the elastic limit of the respective material. In the same way, the low level of stress found on the metaphysis / shaft assembly shows the durability of that combination.

Additionally, during tests conducted by the Laboratoire National d'Essai, France, a modified endurance test was performed, as there were no ISO standards for that type of prostheses ; the tested prostheses resisted 5 million cycles, confirming that the cylindrical insertion absorbed most of the stress reaching that level, without transmitting it to the threading.

METAPHYSIARY
PARTS

RIGHT

LEFT

METAPHYSIARY PARTS

SIZE

NUMBER

SIZE

NUMBER

5

233631

5

233641

7,5

233632

7,5

233642

10

233633

10

233643

11,25

233634

11,25

233644

12,5

233635

12,5

233645

13,75

233636

13,75

233646

15

233637

15

233647

16,25

233638

16,25

233648

17,5

233639

17,5

233649

20

233640

20

233650


DISTAL PART

SIZE AND NUMBER

SIZE AND NUMBER

DISTAL PART

9

201117

13

201121

10

201118

14

203605

11

201119

15

203606

12

201120

.

.


CONE HEADS 12/14

STAINLESS STEEL

ALUMINA

STAINLESS STEEL SKIR

Diam 22,2

Diam 28

Diam 32

Diam 28

Diam 28

 

Col court

.

231391

231386

232967

236270

 

Short neck

231402

231392

231387

232968

232271

 

Long neck

231403

231393

231388

232969

236272

Extra long neck
+ 7 mm

.

231394

.

CONE HEADS

Extra long neck
+ 10 mm

.

231395

.